REFERENCE ELECTRODE

'Reference electrode'
is an electrode which has a stable and well-known electrode potential.
The high stability of the electrode potential is usually reached by employing a redox system with constant (buffered or saturated) concentrations of each participants of the redox reaction.
Reference electrodes are used to measure electrochemical potential.

Contents
Aqueous Reference Electrodes
Nonaqueous Reference Electrodes
References

Aqueous Reference Electrodes


Common reference electrodes and potential with respect to the standard hydrogen electrode:

Standard hydrogen electrode (SHE) (E=0.000V) also known as "normal hydrogen electrode" (NHE)

Saturated calomel electrode (SCE) (E=-0.242V saturated)

Copper-copper(II) sulfate electrode (E=-0.318V)

Silver Chloride Electrode (E=-0.225V saturated)

pH-electrode (in case of pH buffered solutions, see buffer solution)

Palladium-Hydrogen electrode

Nonaqueous Reference Electrodes


While it is convenient to compare between solvents to qualitativley compare systems it is not quantitatively meaningful. Much as pKa are related between solvents, but not the same, so is the case with E°. While the SHE might seem to be a reasonable reference for nonaqueous work as it turns out the platinum is rapidly poisoned by many solvents including acetonitirile causing uncontrolled drifts in potential. Both the SCE and saturated Ag/AgCl are aqueous electrodes based around saturated aqueous solution. While for short periods it may be possible to use such aqueous electrodes as references with nonaqueus solutions the long-term results are not trustworthy. Using aqueous electrodes introduces undefined, variable, and unmeasurable junction potentials to the cell in the form of a liquid-liquid junction as well as different ionic composition between the reference compartment and the rest of the cell. The best argument against using aqueous reference electrodes with nonaqueous systems, as mentioned earlier, is that potentials measured in different solvents are not directly comparable.
'Quasi-Reference Electrode'
Making a quasi-reference electrode (QRE).
i) Inserting a piece of Silver wire into concentrated HCl then allow the wire to dry on a chem-wipe. This forms an insoluble layer of AgCl on the surface of the electrode and gives you a Ag/AgCl wire. Repeat dipping every few months or if the QRE starts to drift.
ii) Obtain a ‘Vycor’ glass frit (4 mm diameter) and glass tubing of similar diameter. Attach ‘Vycor’ glass frit to the glass tubing with heat shrink Teflon tubing.
iii) Rinse then fill the clean glass tube with supporting electrolyte solution and insert Ag/AgCl wire.
iv) The Ferrocene (II/III) couple should lie around 400 mV versus this Ag/AgCl QRE in an acetonitrile solution. This potential will varying up to 200 mV with the specific undefined conditions.
A QRE avoids the issues mentioned above. A QRE with Ferrocene or similar internal standard (Cobaltocene) referenced back to Ferrocene is ideal for nonaqueous work. Ferrocene has been gaining acceptance as the standard reference for nonaqueous work for a number of reasons it is about 60 mV positive of the SHE making it comparable to aqueous systems in a qualitatively sense. The preparation of the QRE electrode is simple allowing a fresh reference to be prepared with each set of experiments. Since QREs are made fresh there is also no concern of improper storage or maintenance of the electrode. QREs are also more affordable than other reference electrodes.

References


Bard, A.J. & L.R. Faulkner, 'Electrochemical Methods: Fundamentals and Applications.' New York: John Wiley & Sons, 2nd Edition, 2000.
D.J.G. Ives & G.J. Janz, 'Reference Electrodes, Theory and Practice.' New York: Academic Press, 1961.
P. Zanello,' Inorganic Electrochemistry: theory, practice and applications.' Cambridge: Royal Society of Chemistry, 2003.