
thumb
'Gas chromatography-mass spectrometry' ('GC/MS') is a method that combines the features of
gas-liquid chromatography and
mass spectrometry to identify different substances within a test sample. Applications of GC/MS include
drug detection,
fire investigation, environmental analysis,
explosives investigation, and identification of unknown samples. GC/MS can also be used in airport security to detect substances in luggage or on human beings. Additionally, it can identify
trace elements in materials that were previously thought to have disintegrated beyond identification.
The GC-MS has been widely heralded as a "
gold standard" for
forensic substance identification because it is used to perform a ''specific test''. A specific test positively identifies the actual presence of a particular substance in a given sample. A ''non-specific test'', however, merely indicates that a substance falls into a category of substances. Although a non-specific test could statistically suggest the identity of the substance, this could lead to
false positive identification.
History
The use of a mass spectrometer as the detector in gas chromatography was developed during the 1950s by Roland Gohlke and Fred McLafferty.
[1][2] These sensitive devices were bulky, fragile, and originally limited to laboratory settings. The development of affordable and
miniaturized computers has helped in the simplification of the use of this instrument, as well as allowed great improvements in the amount of time it takes to analyse a sample. In 1996 the top-of-the-line high-speed GC-MS units completed analysis of fire accelerants in less than 90 seconds, whereas first-generation GC/MS would have required at least 16 minutes. This has led to their widespread adoption in a number of fields.
Instrumentation
The GC-MS is composed of two major building blocks: the
gas chromatograph and the
mass spectrometer. The gas chromatograph utilizes a capillary column and depending on the column's dimensions (length, diameter, film thickness) as well as the phase properties (e.g. (5% (phenyl)polysiloxane) the difference in the chemical properties between different
molecules in a
mixture will separate the molecules as the sample travels the length of the column. The molecules take different amounts of time (called the retention time) to come out of (elute from) the gas chromatograph, and this allows the mass spectrometer downstream to capture, ionize, and detect the molecules separately. The mass spectrometer does this by breaking each molecule into
ionized fragments and detecting these fragments using their mass to charge ratio.

GC-MS schematic
These two components, used together, allow a much finer degree of substance identification than either unit used separately. It is not possible to make an accurate identification of a particular molecule by gas chromatography or mass spectrometry alone. The mass spectrometry process normally requires a very pure sample while gas chromatography using a traditional detector (e.g. Flame Ionization Detector) detect multiple molculears that happen to take the same amount of time to travel through the column (''i.e.'' have the same retention time) which results in two or more molecules to co-elute. Sometimes two different molecules can also have a similar pattern of ionized fragments in a mass spectrometer (mass spectrum). Combining the two processes makes it extremely unlikely that two different molecules will behave in the same way in both a gas chromatograph and a mass spectrometer. Therefore when an identifying mass spectrum appears at a characteristic retention time in a GC-MS analysis, it typically lends to increased certainty that the analyte of interest is in the sample.
Types of Mass Spectrometer Detectors
The most common type of mass spectrometer (MS) associated with a gas chromatograph (GC) is the quadrupole mass spectrometer, sometimes referred to as a Mass Selective Detector. Another relatively common detector is the ion trap mass spectrometer. Additionally one may find a magnetic sector mass spectrometer, however these particular instruments are expensive and bulky and not typically found in throughput laboratories. Other detectors may be encountered such as time of flight (TOF), tandem quadruples (MS/MS), or in the case of an ion trap MS^n where n <= 9.
Analysis
A mass spectrometer is typically utilized in one of two ways: Full Scan or Selective Ion Monitoring (SIM). The typical GC/MS instrument is capable of performing both functions either individually or concomitantly, depending on the setup of the particular instrument.
Full Scan MS
When collecting data in the full scan mode, a target range of mass fragments is determined and inputed into the instrument's method. An example of a typical broad range of mass fragments to monitor would be m/z 50 to m/z 400. The determination of what range to use is largely dictated by what one anticipates in being in the sample while being cognizant of the solvent and other possible interferences. A MS should not be set to look for mass fragments too low or else one may detect air (found as m/z 28 due to nitrogen), carbon dioxide (m/z 44) or other possible interferences. Additionally if one is to use a very large scan range then sensitivity of the instrument is decreased due to performing less scans per second since each scan will have to detect a wide range of mass fragments.
Full Scan is a useful technique when it comes to determining unknown compounds/impurities in a sample. Additionally it provides more information than SIM when it comes to confirming or resolving compounds in a sample. During instrument method development it may be common to first analyze test solutions in full scan mode to determine the retention time and the mass fragment fingerprint before moving to a SIM instrument method.
Selective Ion Monitoring
In Selective Ion Monitoring (SIM) certain specific ion fragments are entered into the instrument method and only those mass fragments will be detected by the mass spectrometer. The benefits of using SIM analysis is that the typical detection limit can be lowered since the instrument is only looking at a handful of fragments (e.g. three fragments) during each scan and therefore more scans can take place each second. Additionally since only the particular mass fragments of interest are being monitored, matrix interferences are typically not as severe as when found in Full Scan mode. To additionally confirm the likelihood of a potentially positive result, it is relatively important to be be sure that the ion ratios of the various mass fragments are comparable to a known reference standard.
Types of Ionization
After the molecules travel the length of the column, pass through the transfer line and enter into the mass spectrometer they are ionized by various methods with typically only one method being used as any given time. Once the sample is fragmented it will then be detected, usually by an electron multiplier diode, which essentially turns the ionized mass fragment into an electrical signal that is then detected.
The ionization technique chosen is independent of using Full Scan or SIM.
Electron Ionization
By far the most common and perhaps standard form of ionization is
electron ionization (EI). The molecules enter into the MS (the source in a quadrupole or the ion trap itself in an ion trap MS) where they are bombarded with free electrons emitted from a filament, not much unlike the filament one would find in a standard light bulb. The electrons bombard the molecules causing a hard ionization that fragments the molecule, and the way in which a molecule fragment is usually typical for all EI techniques.
Chemical Ionization
In
chemical ionization a reagent gas, commonly
methane or
ammonia is introduced into the mass spectrometer. Depending on the technique (positive CI or negative CI) used this reagent gas will interact with the electrons and analyte and cause a 'soft' ionization of the molecule of interest. A softer ionization does not fragment the molecular nearly as much as the harder ionization found with EI. Due to this one of the benefits of using a chemical ionization technique is that a mass fragment closely corresponding to the molecular weight of the analyte of interest is produced.
Positive Chemical Ionization
In Positive Chemical Ionization (PCI) the reagent gas interacts with the target molecule, typically with a proton exchange. This will then produce the [M + H]
+ species in relatively great abundance.
Negative Chemical Ionization
In Negative Chemical Ionization (NCI) the reagent gas will decrease the impact of the free electrons on the target analyte. This decreased energy will typically leave the [M - H]
- fragment in great abundance.
'''The following information is in the process of being updated:'''
The primary goal of instrument analysis is to quantify an amount of substance. This is done by comparing the relative concentrations among the atomic masses in the generated spectrum. Two kinds of analysis are possible, comparative and original. Comparative analysis essentially compares the given spectrum to a spectrum library to see if its characteristics are present for some sample in the library. This is best performed by a
computer because there are a myriad of visual distortions that can take place due to variations in scale. Computers can also simultaneously correlate more data (such as the retention times identified by GC), to more accurately relate certain data.
Another analysis measures the peaks in relation to one another, with the tallest peak receiving 100% of the value, and the others receiving proportionate values, with all values above 3% being accounted for. The parent peak normally indicates the total mass of the unknown compound. This value can then be used to fit to a chemical
formula containing the various
elements assumed to be present in the compound. The
isotope pattern in the spectrum, which is unique for elements having many isotopes, can also be used to identify the various elements present. Once a chemical formula has been matched to the spectrum, the molecular structure and bonding can be identified, and needs to be consistent with a substance with the characteristics recorded by GC/MS. The fitting is normally done automatically by programmes which come with the machine, given a list of the elements which could be present in the sample.
A “full spectrum” analysis considers all the “peaks” within a spectrum. However, selective ion monitoring (SIM) which looks only at a few characteristic peaks associated with a candidate substance, can also be done. This is done on the assumption that at a given retention time, a set of
ions is characteristic of a certain compound. This is a fast and efficient analysis, especially if you have some prior information about a sample or are looking for a specific compound. When the amount of information collected about the ions in a given gas chromatographic peak is reduced, the sensitivity of the analysis goes up. So, SIM analysis allows a smaller quantity of a compound to be detected and measured, but the degree of certainty about the identity of that compound is reduced.
Applications
Environmental Monitoring and Cleanup
GC-MS is becoming the tool of choice for tracking
organic pollutants in the environment. The cost of GC-MS equipment has fallen significantly, and the reliability has increased at the same time, which has contributed to its increased adoption in environmental studies as cost is always a major consideration in this kind of work. There are some compounds for which GC-MS is not sufficiently sensitive, including certain pesticides and herbicides, but for most organic analysis of environmental samples, including many major classes of pesticides, it is very sensitive and effective.
Criminal Forensics
GC-MS can analyze the particles from a human body in order to help link a criminal to a
crime. The analysis of
fire debris using GC-MS is well established, and there is even an established American Society for Testing Materials (ASTM) standard for fire debris analysis.
Law Enforcement
GC-MS is increasingly used for detection of illegal narcotic, and may eventually supplant drug-sniffing dogs.
Security
A post-
September 11 development,
explosive detection systems have become a part of all
US airports. These systems run on a host of technologies, many of them based on GC-MS. There are only three manufacturers certified by the
FAA to provide these systems (citation needed), one of which is Thermo Detection (formerly Thermedics), which produces the
EGIS, a GC-MS-based line of explosives detectors. The other two manufacturers are Barringer Technologies, now owned by Smith's Detection Systems and Ion Track Instruments, part of General Electric Infrastructure Security Systems.
Food, Beverage and Perfume Analysis
Foods and
beverages contain numerous
aroma compounds, some naturally present in the raw materials and some forming during processing. GC-MS is extensively used for the analysis of these compounds which include
esters,
fatty acids,
alcohols,
aldehydes,
terpenes etc. It is also used to detect and measure contaminants from spoilage or
adulteration which may be harmful and which is often controlled by governmental agencies, for example
pesticides.
Astrochemistry
Several GC-MS have left earth. Two were brought to
mars by the
Viking program.
[3] Venera 11 and 12 and
Pioneer Venus analysed the atmosphere of
venus with GC-MS.
[4] The
Huygens probe of the
Cassini-Huygens mission landed one GC-MS on
Saturn's largest moon,
Titan.
[5] The material in the
comet 67P/Churyumov-Gerasimenko will be analysed by the
Rosetta mission with a chiral GC-MS in 2014.
[6]
Medicine
In combination with
isotopic labeling of metabolic compounds, the GC-MS is used for determining
metabolic activity. Most applications are based on the use of
13C as the labeling and the measurement of
13C/
12C ratios with an 'isotope ratio mass spectrometer' ('IRMS'); an MS with a detector designed to measure a few select ions and return values as ratios.
See also
★
mass spectrometer
★
gas chromatograph
References
1. Gohlke, R. S., Time-of-flight mass spectrometry and gas-liquid partition chromatography. ''Anal. Chem.'' '1959', ''31'', 535-41
2. Gohlke, R. S.; McLafferty, F. W., Early gas chromatography/mass spectrometry.'' J. Am. Soc. Mass Spectrom.'' '1993', ''4'', (5), 367-371.
3. The Development of the Viking GCMS
4. Chemical composition of the atmosphere of Venus, V. A. Krasnopolsky, V. A. Parshev, , , Nature, 1981
5. The abundances of constituents of Titan’s atmosphere from the GCMS instrument on the Huygens probe, H. B. Niemann, S. K. Atreya, S. J. Bauer, G. R. Carignan, J. E. Demick, R. L. Frost, D. Gautier, J. A. Haberman, D. N. Harpold, D. M. Hunten, G. Israel, J. I. Lunine, W. T. Kasprzak, T. C. Owen, M. Paulkovich, F. Raulin, E. Raaen, S. H. Way, , , Nature, 2005
6. COSAC onboard Rosetta: A bioastronomy experiment for the short-period comet 67P/Churyumov-Gerasimenko, Goesmann F, Rosenbauer H, Roll R, Bohnhardt H, , , Astrobiology, 2005
★ Eiceman, G.A. (2000). Gas Chromatography. In R.A. Meyers (Ed.), ''Encyclopedia of Analytical Chemistry: Applications, Theory, and Instrumentation'', pp. 10627. Chichester: Wiley. ISBN 0-471-97670-9
★ Giannelli, Paul C. and Imwinkelried, Edward J. (1999). Drug Identification: Gas Chromatography. In ''Scientific Evidence'' '2', pp. 362. Charlottesville: Lexis Law Publishing. ISBN 0-327-04985-5.
External links
★
★
GCMS Tutorial
★
Gas Chromatography-Mass Spectroscopy Background
★
Introduction to Mass Spectrometry